2A1-L06 A surgical robot with heart-surface-motion synchronization mechanism for myoblast cell sheet transplantation(Surgical Robotics and Mechatronics (1))
Kangyi XURyoichi NAKAMURA
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Pages _2A1-L06_1-_2A1-L06_4

Details
Abstract
Myoblast cell sheets are taking into the clinical treatment of heart disorders. In our study, we proposed a surgical robot system, includes a robot system with the characteristics of Double RCM (remote center of motion) mechanism, which can cancel and compensation the beating heart surface movement for the myoblast cell sheet transplantation and two endoscopic cameras to capture heart surface motion without obstruct the transplantation of the sheets. The linear and rotation motion of the robot system are totally isolated, realized by employed a Double RCM mechanism. An experiment is conducted to evaluate the tracking accuracy of our robot system in tracking an approximate -period moving target. The tracking data is updating by Polaris system in 30Hz. From the experiment, we got the results that the linear and error is 4.67±3.73 mm and the rotation tracking error is 2.86±5.50 degree.