Hyomen Kagaku
Online ISSN : 1881-4743
Print ISSN : 0388-5321
ISSN-L : 0388-5321
Note
Minute Current Detection by Scanning Probe Microscopy under High Humidity Atmospheres
Kazunori ANDOHiromi KURAMOCHIHiroshi YOKOYAMA
Author information
JOURNAL FREE ACCESS

2003 Volume 24 Issue 9 Pages 586-589

Details
Abstract
Detection of pA order minute current by scanning probe microscopy encounters some problems under high humidity atmospheres, such as the increased background level and the poor reproducibility, because of the surface leak current. We were able to overcome these problems by using hermetic sealing and keeping the whole electronics equipment in a humidity-controlled atomic force microscope system. Detection of minute current of sub-pA level has been achieved even under high humidity atmospheres.