The Journal of Toxicological Sciences
Online ISSN : 1880-3989
Print ISSN : 0388-1350
ISSN-L : 0388-1350
Letter
Quantification of molecules in 1H-NMR metabolomics with formate as a concentration standard
Itiro AndoTakuo HiroseTadashi NemotoKazuhito TotsuneYutaka ImaiKazuhisa TakeuchiMasako Fujiwara
著者情報
ジャーナル フリー

2010 年 35 巻 2 号 p. 253-256

詳細
抄録

Quantitative analysis of metabolites is important in 1H-nuclear magnetic resonance (NMR)-based metabolomics of plasma. Human plasma contains a high density of proteins which heavily adsorb the commonly-used standard compound of sodium 3-(trimethylsilyl) propionate 2, 2, 3, 3-d4 (TSP). We have evaluated calcium formate as an alternative standard in 1D single-pulse 1H-NMR spectra to quantify plasma metabolites. Formate did not interact with either plasma metabolites or proteins under adequate conditions. Linear relations between the signal intensities and the added formate have been demonstrated in 1H spectra. The quantifications of glucose and creatinine by this method have shown good accordance with biochemical analysis. Calcium formate is applicable as a concentration standard to NMR metabolomics of plasma.

著者関連情報
© 2010 The Japanese Society of Toxicology
前の記事 次の記事
feedback
Top