Journal of UOEH
Online ISSN : 2187-2864
Print ISSN : 0387-821X
ISSN-L : 0387-821X
Thermal Distribution in the Agar Phantom by a New Intracavitary RF Applicator for Prostate Gland
Hisato INATOMIAtsuo SUGITAHiromi TERASHIMATakao YOSHIURANaoki KUNUGITAToshiyuki NORIMURATakehiko TSUCHIYA
Author information
JOURNAL FREE ACCESS

1992 Volume 14 Issue 1 Pages 39-45

Details
Abstract
Some studies suggest that the prostate gland is one of the most difficult portions of the body to heat up even by radio frequency (RF) capacitive heating, which is well established as a useful means for deep regional hyperthermia because of its non-uniform power absorption. We have developed a new type of intracavitary applicator for RF capacitive heating which can be connected to an 8MHz RF heating device (Thermotron-RF8). The applicator is a balloon-type catheter which is filled with electrolite, such as physiological saline, 3% saline or 3% ferrous sulfate solution, and then put between a pair of regular RF capacitive heating electrodes. The heating characteristics of the applicator were examined by using an agar phantom that is electrically equivalent to muscles. When the applicator was connected to the electrode of the RF generator with a copper cord, the balloon filled with electrolite operated as an inner electrode. The effective hot area around the balloon of the catheter was observed, and the size of the hot area was changed by the kinds of electrolites used and their concentration. With this adaptive technique, it is possible to maintain an adequate thermal distribution at a tumor in the uterus, urinary bladder and rectum, as well as the prostate gland.
Content from these authors
© 1992 The University of Occupational and Environmental Health, Japan
Previous article Next article
feedback
Top