Journal of the Visualization Society of Japan
Online ISSN : 1884-037X
Print ISSN : 0916-4731
ISSN-L : 0916-4731
Performance Characteristics of a Nasal CPAP Device
Kenji HOSOIMasaaki KAWAHASHISou MATSUMOTOHiroyuki HIRAHARA
Author information
JOURNAL FREE ACCESS

2006 Volume 26 Issue Supplement2 Pages 307-310

Details
Abstract
Artificial respiration techniques for patients with respiratory disorder are important as a basic life support treatments for the patients. The nasal CPAP (Continuous Positive Airway Pressure) is a typical clinical treatment for infants keeping self respiration. Some devices for the nasal CPAP have been used in clinical procedure. However, in order to reduce the load for patients in the treatments, the downsizing and improvement of fitting of the devices are required. A new type device of nasal CPAP has been developed by the authors, but its performance in MAP (Mean Airway Pressure) variation did not meet the condition for clinical use. In this report, the performance of a nasal CPAP device improved has been investigated and it has been clarified that the performance was meet to clinical use.
Content from these authors
© The Visualization Society of Japan
Previous article Next article
feedback