Journal of the Vacuum Society of Japan
Online ISSN : 1882-4749
Print ISSN : 1882-2398
ISSN-L : 1882-2398
Article
Development of an Apparatus for High-Resolution Auger Photoelectron Coincidence Spectroscopy (APECS) and Electron Ion Coincidence (EICO) Spectroscopy
Takuhiro KAKIUCHIShogo HASHIMOTONarihiko FUJITAKazuhiko MASEMasatoshi TANAKAMakoto OKUSAWA
Author information
JOURNAL FREE ACCESS

2008 Volume 51 Issue 11 Pages 749-757

Details
Abstract
  We have developed an electron electron ion coincidence (EEICO) apparatus for high-resolution Auger photoelectron coincidence spectroscopy (APECS) and electron ion coincidence (EICO) spectroscopy. It consists of a coaxially symmetric mirror electron energy analyzer (ASMA), a miniature double-pass cylindrical mirror electron energy analyzer (DP-CMA), a miniature time-of-flight ion mass spectrometer (TOF-MS), a magnetic shield, an xyz stage, a tilt-adjustment mechanism, and a conflat flange with an outer diameter of 203 mm. A sample surface was irradiated by synchrotron radiation, and emitted electrons were energy-analyzed and detected by the ASMA and the DP-CMA, while desorbed ions were mass-analyzed and detected by the TOF-MS. The performance of the new EEICO analyzer was evaluated by measuring Si 2p photoelectron spectra of clean Si(001)-2×1 and Si(111)-7×7, and by measuring Si-L23VV-Si-2p Auger photoelectron coincidence spectra (Si-L23VV-Si-2p APECS) of clean Si(001)-2×1.
Content from these authors
© 2008 The Vacuum Society of Japan
Previous article Next article
feedback