Journal of the Vacuum Society of Japan
Online ISSN : 1882-4749
Print ISSN : 1882-2398
ISSN-L : 1882-2398
Review
Outgassing Measurement for Vacuum Engineering
Sakae INAYOSHI
Author information
JOURNAL FREE ACCESS

2015 Volume 58 Issue 2 Pages 57-62

Details
Abstract
  In this article, two common methods to evaluate the outgassing properties of vacuum system materials, thermal desorption spectroscopy (TDS) and the method of switching between two pumping paths (SPP), are discussed. TDS can evaluate the outgassing properties of small samples, while SPP method can measure low outgassing rate, such as the case after baking. Ion gauges and/or quadrupole mass spectrometers (QMS) consist of the TDS and SPP setups. In these cases, QMS are typically calibrated using a standard conductance element (SCE). The calibration using the SCE enables the direct conversion of the QMS ion current into the gas flow rate.
Content from these authors
© 2015 The Vacuum Society of Japan
Previous article Next article
feedback
Top