2010 Volume 45 Issue 2 Pages 137-139
We imeplemented the method of time-delayed feedback control to a commercial dynamic-mode atomic force microscope for suppressing chaotic oscillation caused by the nonlinear tip-sample interaction force. Cantilever oscillation was successfully stabilized by applying our prototype time-delayed feedback controller with the magnetic cantilever excitation method.