ライフサポート
Online ISSN : 1884-5827
Print ISSN : 1341-9455
ISSN-L : 1341-9455
研究論文
体重計による尿流測定のためのアルゴリズム開発
野呂 剛史金 主賢藤田 徹也中島 一樹
著者情報
キーワード: Uroflowmetry, Bath scale, Smoothing
ジャーナル フリー

2019 年 31 巻 2 号 p. 64-70

詳細
抄録

Acquisition of daily urination information leads to accurate diagnosis and treatment of dysuria. Conventional uroflowmeters have some problems about installation cost, sanitation management and measurement condition. As a method to solve them, uroflowmetry by the bath scale was proposed. In this study, flow rate change was calculated from weight change and smoothed by moving average to remove effect of body sway. First, pseudo-urination was performed on the bath scale and an optimum number for smoothing was determined. Next, urination was performed on the bath scale, and it was measured simultaneously with a uroflowmeter. The accuracy of 5 measurements was examined with 7 subjects. As a result, the bell-shaped urinary flow patterns followed to the patterns of uroflowmeter. The estimated voided volume (VVest) and maximum flow rate (Qmax) distributed within ±20% of the reference.

著者関連情報
© 2019, ライフサポート学会
前の記事 次の記事
feedback
Top