レーザー研究
Online ISSN : 1349-6603
Print ISSN : 0387-0200
ISSN-L : 0387-0200
レーザー解説
テラヘルツ波絶対パワーの精密測定
飯田 仁志木下 基雨宮 邦招
著者情報
ジャーナル フリー

2019 年 47 巻 1 号 p. 37-

詳細
抄録
To realize reliable terahertz sensing applications, accurate calibration of sources and detectors is essential. Since power is a basic physical quantity, its absolute measurement is important to evaluate the performance of the sensors. This paper reviews recent international trends in terahertz power measurements and standards. A highly sensitive calorimeter is introduced to measure absolute terahertz power below microwatt levels at room temperature. The calorimeter measures the absolute terahertz power by substituting dc power using an isothermal calorimetric technique. A power meter is calibrated using the calorimeter as a reference standard. This calorimeter is expected to realize the International System of Units (SI)-traceable measurements of terahertz power.
著者関連情報
© 2019 一般社団法人 レーザー学会
前の記事 次の記事
feedback
Top