Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
REGULAR PAPERS
Precise Measurement of Nitrogen Isotopic Composition Using a Quadrupole Mass Spectrometer
Ko HashizumeNaoji Sugiura
Author information
JOURNAL FREE ACCESS

1990 Volume 38 Issue 6 Pages 269-286

Details
Abstract
A system to precisely analyze the isotope of nanogram-order nitrogen was constructed using a quadrupole mass-spectrometer. Neon and argon isotopes can also be measured by the system. The reproducibility of the nitrogen isotopic ratio is better than 3‰. The minimum amount of standard air nitrogen required for precise isotopic determination is 3 ng. The system is oriented to measure extra-terrestrial samples with low nitrogen concentration. The detail of the stepwise combustion method to separate a trace amount of indigenous nitrogen from the terrestrial nitrogen is also reported.
Content from these authors
© 1990 by The Mass Spectrometry Society of Japan
Next article
feedback
Top