Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
報文
Effect of Ammonia Addition to the Mobile Phase on Atmospheric Pressure Chemical Ionization Mass Spectrometry of Methyl Hydroperoxyoleates
Yoshiyuki MOCHIDA, Yukio YOKOYAMA, Shigeo NAKAMURA
著者情報
ジャーナル フリー

1998 年 46 巻 3 号 p. 246-249

詳細
抄録
The method for observing molecular related ions of methyl hydroperoxyoleates (OMH) by atmospheric pressure chemical ionization (APCI) mass spectrometry (MS) was investigated. Since thermally labile OMH is decomposed in the APCI nebulizer and/or desolvation chamber heated to a high temperature, it is difficult to observe significant peaks of their protonated molecules. The addition of ammonia water to the mobile phase (ethanol) was quite effective to suppress the thermal decomposition of OMH. As the result the mass spectra exhibited the intense peaks of ammonium adduct ions of OMH, [M+NH4]+, at m/z 346. The nebulizer temperature at 170°C and the drift voltage at 170 V were adequate for observing the molecular ion species of OMH.
著者関連情報
© 1998 by The Mass Spectrometry Society of Japan
前の記事 次の記事
feedback
Top