Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
速報
Optimization of Matrix and Amount of Ammonium Chloride Additive for Effective Ionization of Neutral Oligosaccharides as Chloride Ion Adducts in Negative-Mode MALDI-TOF Mass Spectrometry
Hiroaki SUZUKITohru YAMAGAKIKazuo TACHIBANA
著者情報
ジャーナル フリー

2005 年 53 巻 4 号 p. 227-229

詳細
抄録
The combination of harmine and ammonium chloride is a useful matrix in detecting the chloride ion adducts of neutral oligosaccharides in negative-ion MALDI-TOF mass spectra. Sandwich method is the best way to prepare the mixed crystals of matrix, analyte, and additive on the target. Changing the molar ratio of ammonium chloride additive from 0 to five times the amount of harmine matrix, the relative intensity of the chloride ion adducts of the analyte molecules to the deprotonated matrix molecular ion ( [Analyte+Cl]-/ [Matrix-H]-) increases logarithmically. A ratio of several times the amount of additive to matrix works most effectively in the formation and detection of the chloride ion adducts of the neutral oligosaccharides as opposed to an amount of additive that is on the same order as the analytes.
著者関連情報
© 2005 by The Mass Spectrometry Society of Japan
前の記事 次の記事
feedback
Top