Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
総合論文
Nano-ESIプローブの開発と高分解能測定法およびMIKEスペクトル法を用いた有機イオンの分解機構に関する研究
高橋 豊
著者情報
ジャーナル フリー

2005 年 53 巻 6 号 p. 319-328

詳細
抄録
A new technique is described for accurate mass measurement by using multiple sprayers nano electrospray ionization mass spectrometry (nano-ESI/MS). A nano-scale high-performance liquid chromatography (nano-LC) was combined with a magnetic sector instrument. A metal-coated glass capillary was used for the introduction of the reference compound solution, and a metal-coated fused silica capillary was used to connect the nano-LC column to ionization chamber. By shifting the position of the each sprayers to the sampling orifice, the spectra were obtained for the sample compounds eluted from the column and the reference one. Several standard compounds were examined by using this technique and satisfactory accurate masses were obtained. It was not problems for the differences between the ionization efficiency of the sample and the reference compounds. Studies on the unimolecular metastable ion decomposition of several organic compounds are also described by using high-resolution mass spectrometry, MIKE (mass-analyzed ion kinetic energy) spectrometry and D-labeling technique. The m/z 45 ions generated from 2-methoxyethylamine (CH3OCH2CH2NH2) consist of two isobaric ions, methoxymethyl ion, CH3OCH2+, and C2H7N+. The m/z 69 ions generated from methyl trifluoroacetoacetate (CF3COCH2COOCH3) and 1,1,1,5,5,5-hexafluoro-2,4-pentanedione (CF3COCH2COCF3) also consist of two isobaric ions, trifluoromethyl ion, CF3+, and protonated carbon suboxide, OCCH+CO.
著者関連情報
© 2005 日本質量分析学会
前の記事 次の記事
feedback
Top