An Automatic Ratio Readout System for a Double Collection Mass Spectrometer
EDWARD M. BEAVER
著者情報
ジャーナル フリー

1973 年 21 巻 1 号 p. 37-44

詳細
抄録
An improved isotope ratio readout system for a six-inch ratio mass spectrometer has been developed that offers the advantages of automatic operation. precision equivalent to or somewhat better than other systems, and low cost. A pair of solid-state vibrating reed electrometers are used to measure the ion currents from two isotopes. These are interfaced to a frequency counter, operated in ratio mode, through two voltage-to-frequency converters. A digital programer is used for automatic operation.