医用画像情報学会雑誌
Online ISSN : 1880-4977
Print ISSN : 0910-1543
ISSN-L : 0910-1543
臨床MR画像を対象としたSNR測定法の検討
—同一関心領域法におけるSD mapの補正法の開発—
林 則夫峰広 香織作田 啓太宮地 利明真田 茂鈴木 正行松浦 幸広河原 和博山本 友行
著者情報
ジャーナル フリー

2009 年 26 巻 1 号 p. 8-12

詳細
抄録
Signal-to-Noise ratio (SNR) is an important quantity factor used to describe the performance of a magnetic resonance (MR) imaging. However, it is difficult to measure accurately the SNR in clinical parallel MR images. This study is to develop a SNR measurement method using a modified identical region of interest (ROI) technique in clinical parallel MR images. Phantom experiments were performed on a 1.5 and 3.0 Tesla MRI machine to compare three different methods of SNR measurements. We also performed SNR measurements in clinical cases. The method is consisted of the following four steps: 1) Calculation of standard deviation (SD) map 2) Edge detection and removal on SD map 3) Interpolation of the removed edge region 4) Calculation of SNR map using the interpolated SD map. Our developed method were approximately consistent with those of other SNR measurement methods in both phantom and clinical images. In addition, the results of our developed method could measure the SNR in the edge region in parallel MR images. Our method makes it possible to simply and accurately analyze SNR of the clinical parallel image with a single scan.
著者関連情報
© 2009 医用画像情報学会
前の記事 次の記事
feedback