歯科用CBCTにおける回転角度の違いが画質に及ぼす影響:物理評価による検討
伊藤 実咲小野寺 崇小嶋 郁穂千田 浩一
著者情報
ジャーナル フリー

2022 年 39 巻 2 号 p. 38-42

詳細
抄録

Dental cone beam computed tomography (CBCT) is now widely available withboth 180° and 360°imaging. Radiation dose and motion artifacts are lower for 180° than 360° imaging. Therefore, 180° imaging may be useful in children who are more sensitive toradiation exposure and have difficulty in controlling their body movements. Thepurpose of this study was to clarify the effect of rotation angle on imagequality from modulation transfer function (MTF) and noise power spectrum (NPS) and to determine the optimal imaging tube voltage byobtaining the tube voltage characteristics of FPD. As a result, resolution wasnot affected by rotation angle. NPS was higher in 180°imaging than in 360° imaging. System performance (SP) function was lower in 180° imaging than in 360° imaging, withoutchanging the frequency characteristics. The higher the tube voltage, the higherthe value of SP function and the lower the radiation dose. The optimal tubevoltage for dental CBCT will be 90 kV.

著者関連情報