Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Notes
Intersite Reliability of Diffusion Tensor Imaging on Two 3T Scanners
Koji KAMAGATA Keigo SHIMOJIMasaaki HORIAkira NISHIKORIKohei TSURUTAMariko YOSHIDAKouhei KAMIYARyusuke IRIEMichimasa SUZUKIShinsuke KYOGOKUYuriko SUZUKINoriko SATOShigeki AOKI
著者情報
ジャーナル オープンアクセス

2015 年 14 巻 3 号 p. 227-233

詳細
抄録
We report the intersite scan reliability of diffusion tensor imaging (DTI) parameters using identical 3T scanners and acquisition protocols at 2 sites. Voxel-based analysis revealed several regions with significant intersite differences. The intersite reliability of DTI measures showed coefficients of variation below 4% in tract-specific analysis (TSA) and below 6% in atlas-based analysis. Given the excellent reliability of TSA, our results suggest it as a promising and useful tool for multicenter DTI studies.
著者関連情報
© 2015 by Japanese Society for Magnetic Resonance in Medicine
前の記事 次の記事
feedback
Top