Quantitative T2 Mapping of Knee Cartilage: Comparison between the Synthetic MR Imaging and the CPMG Sequence
Seung Hyun LeeYoung Han LeeHo-Taek SongJin-Suck Suh
Author information
JOURNAL OPEN ACCESS

2018 Volume 17 Issue 4 Pages 344-349

Details
Abstract

The purpose was to evaluate the feasibility of quantitative MRI T2 mapping based on the quantitative MRI (QRAPMASTER) sequence for the quantitative assessment of knee cartilage. The T2 values from the phantom study showed excellent correlation between the two techniques (r2 = 0.998). The cartilage T2 values exhibited strong correlations (r2 = 0.867–0.982). Quantitative MRI (qMRI) T2 mapping can be used as an alternative to multi-echo T2 mapping, with relatively short scan time.