Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Utility of Stack-of-stars Acquisition for Hepatobiliary Phase Imaging without Breath-holding
Shintaro IchikawaUtaroh MotosugiMarie-Luise KromreyDaiki TamadaTetsuya WakayamaKang WangTy A CashenAli ErsozHiroshi Onishi
著者情報
ジャーナル オープンアクセス

2020 年 19 巻 2 号 p. 99-107

詳細
抄録

Purpose: Post-contrast liver magnetic resonance imaging is typically performed with breath-hold 3D gradient echo sequences. However, breath-holding for >10 s is difficult for some patients. In this study, we compared the quality of hepatobiliary phase (HBP) imaging without breath-holding using the prototype pulse sequences stack-of-stars liver acquisition with volume acceleration (LAVA) (LAVA Star) with or without navigator echoes (LAVA Starnavi+ and LAVA Starnavi−) and Cartesian LAVA with navigator echoes (Cartesian LAVAnavi+).

Methods: Seventy-two patients were included in this single-center, retrospective, cross-sectional study. HBP imaging using the three LAVA sequences (Cartesian LAVAnavi+, LAVA Starnavi−, and LAVA Starnavi+) without breath-holding was performed for all patients using a 3T magnetic resonance system. Two independent radiologists qualitatively analyzed (overall image quality, liver edge sharpness, hepatic vein clarity, streak artifacts, and respiratory motion/pulsation artifacts) HBP images taken by the three sequences using a five-point scale. Quantitative evaluations were also performed by calculating the liver-to-spleen, -lesion, and -portal vein (PV) signal intensity ratios. The results were compared between the three sequences using the Friedman test.

Results: LAVA Starnavi+ showed the best image quality and hepatic vein clarity (P < 0.0001). LAVA Starnavi− showed the lowest image quality (P < 0.0001–0.0106). LAVA Starnavi+ images showed fewer streak artifacts than LAVA Starnavi− images (P < 0.0001), while Cartesian LAVAnavi+ images showed no streak artifacts. Cartesian LAVAnavi+ images showed stronger respiratory motion/pulsation artifacts than the others (P < 0.0001). LAVA Starnavi− images showed the highest liver-to-spleen ratios (P < 0.0001–0.0005). Cartesian LAVAnavi+ images showed the lowest liver-to-lesion and -PV ratios (P < 0.0001–0.0108).

Conclusion: In terms of image quality, the combination of stack-of-stars acquisition and navigator echoes is the best for HBP imaging without breath-holding.

著者関連情報
© 2020 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top