Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Notes
Evaluation of Thin-slice Coronal Single-shot Turbo Spin-echo Diffusion-weighted Imaging of the Hand: A Comparison with Conventional Echo-planar Diffusion-weighted Imaging
Masaki OgawaMotoo NakagawaNobuyuki AraiHirohito KanShota OhbaShunsuke ShibataHiroyuki MakiYuta Shibamoto
著者情報
ジャーナル オープンアクセス

2020 年 19 巻 4 号 p. 382-388

詳細
抄録

We prospectively evaluated thin-slice coronal turbo spin-echo (TSE) diffusion-weighted imaging (DWI) covering a larger area with the recently-developed techniques on a 3T MRI scanner, compared with echo-planar imaging (EPI)-DWI in patients undergoing routine hand MRI. Visual score assessment and apparent diffusion coefficient (ADC) measurement were performed for patients with suspected hand tumors. TSE-DWI was superior to EPI-DWI, with less image distortion. The visual score and ADC comparison assessments proved that the image noise of TSE-DWI was acceptable.

著者関連情報
© 2019 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.