Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Notes
MR Microscopy of the Articular Cartilage with a 1.0T Permanent Magnet Portable MR System: Preliminary Results
Hiroshi YOSHIOKATomoyuki HAISHITakaaki UEMATSUYoshimasa MATSUDAYuji ITAIKatsumi KOSE
著者情報
ジャーナル オープンアクセス

2003 年 2 巻 1 号 p. 51-55

詳細
抄録
The present study was designed to evaluate the use of a 1.0T portable permanent magnet MR system in obtaining microscopic MR images of the hyaline cartilage in vitro. A clear laminar appearance was demonstrated with this system. In addition, it was possible to demonstrate cartilage surface irregularity, a decrease in cartilage thickness, and T2 alteration by proteoglycan depletion following up to 12 hours of trypsin treatment. In summary, the portable MR system is useful for investigating cartilage in vitro.
著者関連情報
© 2003 by Japanese Society for Magnetic Resonance in Medicine