Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Investigating Signal Loss due to a Carotid Artery Stent in 3D-TOF-MRA
Hiroshi KatoNorio OotaniKentaro AbiruMika Okahara
Author information
JOURNAL OPEN ACCESS

2021 Volume 20 Issue 3 Pages 303-311

Details
Abstract

Purpose: In this study, we investigated the factors of signal loss out because of the presence of a stent and optimized imaging parameters for improvement in depiction ability.

Methods: We investigated the relationship between the stent type and magnetic susceptibility effect by measuring the signal value between the inside and outside of the stent with different Bw and TE for two different kinds of stents respectively. Similarly, flip angles were changed for two different kinds of stents respectively to the signal intensity between the inside and the outside of the stent was measured, in which examine the relationship between the stent type and the Ernst angles in RF-shielding effect. The conventional imaging parameters and the optimum imaging parameters for each stent obtained from the result of the phantom experiment were examined. Optimized 3D time-of-flight MR angiography (3D-TOF-MRA) was performed and compared with conventional 3D-TOF-MRA and computed tomography angiography (CTA).

Results: The influence of the magnetic susceptibility effect is small in the central part of Carotid Wallstent and in PRECISE, and large in the Carotid Wallstent at the both ends. The influence of RF-shielding effect was large at PRECISE, where the Ernst angle was greatly shifted while the effect is no longer influenced at Carotid Wallstent. Both Carotid Wallstent and PRECISE made imaging capability improved by optimizing the imaging parameters.

Conclusion: During clinical imaging of patients post-carotid artery stenting (CAS) using our protocol, the ability to visualize blood vessels was improved. Thus, we demonstrated that the ability of 3D-TOF-MRA post-CAS was improved via optimizing imaging parameters.

Content from these authors
© 2020 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/