Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Note
Accuracy of MR Temperature Measurement Based on Chemical Shift Change for Radiofrequency Ablation Using Hook-shaped Electrodes
Shigeo OKUDAKagayaki KURODAOsamu KAINUMAKoichi OSHIOHirokazu FUJIWARASachio KURIBAYASHI
Author information
JOURNAL OPEN ACCESS

2004 Volume 3 Issue 2 Pages 95-100

Details
Abstract
The purpose of the current study is to evaluate the accuracy of MR thermometry for radiofrequency ablation (RFA) with hook-shaped electrodes. The objects were eight extracted bovine livers. The chemical shift change was calculated from MR images acquired with a spoiled gradient echo sequence and compared with the temperature directly measured with a thermocouple. Linear regression was established between them with a coefficient of -0.0110±0.0007 ppm/°C and errors were calculated as -0.50±7.50°C. MR thermometry is capable of monitoring temperature for RFA.
Content from these authors
© 2004 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback
Top