Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Reviews
High-speed Imaging at 3 Tesla: A Technical and Clinical Review with an Emphasis on Whole-brain 3D Imaging
Shinji NAGANAWAHisashi KAWAIHiroshi FUKATSUTakeo ISHIGAKITomomi KOMADAKatsuya MARUYAMAOsamu TAKIZAWA
Author information
JOURNAL OPEN ACCESS

2004 Volume 3 Issue 4 Pages 177-187

Details
Abstract

Improvements to the inherently high specific-absorption rate (SAR) of high-speed imaging at 3T are necessary in order to render this method clinically feasible. Various efforts have been undertaken to improve the associated hardware and software. In this review, we focus on whole-brain isotropic 3D imaging with a turbo spin-echo sequence with variable flip-angle echo trains (3D-TSE-VFL) and present its technical and clinical features. This sequence can be used to acquire images of various contrasts including T2-weighted, fat-suppressed T2-weighted, fluid-attenuated inversion recovery (FLAIR), fat-suppressed FLAIR, and STIR (short tau inversion recovery). Various aspects of 3D-TSE-VFL are discussed, including CSF (cerebrospinal fluid) and metal artifacts, STIR contrast, small-part visualization other than brain, and the possibility of serial subtraction. Some images from clinical cases are presented.

Content from these authors
© 2004 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback