Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Notes
Averaging Keyhole Pulse Sequence with Presaturation Pulses and EXORCYCLE Phase Cycling for Dynamic Contrast-Enhanced MRI
Hideto KURIBAYASHIDaniel P. BRADLEYDavid R. CHECKLEYPhilip L. WORTHINGTONJean J. TESSIER
著者情報
ジャーナル オープンアクセス

2004 年 3 巻 4 号 p. 207-210

詳細
抄録
The purpose of this study was to design a keyhole pulse sequence for quantitative 2D dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) based on a spoiled gradient echo T1-weighted acquisition. Saturation recovery was applied to achieve a linear correlation between signal intensity and contrast agent concentration in an arterial input function (AIF) while simultaneously removing time-of-flight effect. To remove ghosting artifacts arising from incomplete presaturation, EXORCYCLE phase cycling with averaging was applied to the pulse sequence. RF spoiling by radiofrequency switching with the synthesizer can be combined with EXORCYCLE phase cycling. Images affected by the large difference in signal intensity before and after contrast agent administration with the keyhole technique were improved by interleaving of peripheral lines of k-space with groups of central lines. Both peripheral and central lines were renewed during the dynamic scan. AIFs were obtained from the rat abdominal aorta with this keyhole sequence.
著者関連情報
© 2004 by Japanese Society for Magnetic Resonance in Medicine
前の記事 次の記事