Simultaneous Three-dimensional Visualization of the Intra-parotid Facial Nerve and Parotid Duct using a Three-dimensional Reversed FISP Sequence with Diffusion Weighting
Shinji NAGANAWAShunichi ISHIHARAHiroko SATAKEHisashi KAWAIMichihiko SONETsutomu NAKASHIMA
著者情報
ジャーナル オープンアクセス

2010 年 9 巻 3 号 p. 153-158

詳細
抄録

We visualized the intraparotid facial nerve and parotid duct at 3 tesla using 3-dimensional reversed fast imaging with steady-state precession (FSIP) (3D-PSIF) with diffusion weighting. Excellent fat suppression by water-selective excitation, sufficient T2-weighting of 3D-PSIF, vessel suppression by diffusion weighting, and high spatial resolution allowed the simultaneous visualization. We also present volumetric representation of the facial nerve and duct.

著者関連情報
© 2010 by Japanese Society for Magnetic Resonance in Medicine