Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182

This article has now been updated. Please use the final version.

A Pitfall of the Volume Rendering Method with 3D Time-of-Flight MRA: A Case of a Branching Vessel at the Aneurysm Neck
Masami GOTOAkira KUNIMATSUMasaaki SHOJIMAOsamu ABEShigeki AOKINaoto HAYASHIHarushi MORIKenji INOKeiichi YANONobuhito SAITOKuni OHTOMO
Author information
JOURNAL FREE ACCESS Advance online publication

Article ID: 2012-0043

Details
Abstract
We present a case in which the origin of the branching vessel at the aneurysm neck was observed at the wrong place on the volume rendering method (VR) with 3D time-of-flight MRA (3D-TOF-MRA) with 3-Tesla MR system. In 3D-TOF-MRA, it is often difficult to observe the origin of the branching vessel, but it is unusual for it to be observed in the wrong place. In the planning of interventional treatment and surgical procedures, false recognition, as in the unique case in the present report, is a serious problem. Decisions based only on VR with 3D-TOF-MRA can be a cause of suboptimal selection in clinical treatment.
Content from these authors
© 2013 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
feedback
Top