Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182

This article has now been updated. Please use the final version.

The Technical and Clinical Features of 3D-FLAIR in Neuroimaging
Shinji NAGANAWA
Author information
JOURNAL FREE ACCESS Advance online publication

Article ID: 2014-0132

Details
Abstract

In clinical MR neuroimaging, 3D fluid-attenuated inversion recovery (3D-FLAIR) with a variable-flip-angle turbo spin echo sequence is becoming popular. There are more than 100 reports regarding 3D-FLAIR in the PubMed database. In this article, the technical and clinical features of 3D-FLAIR for neuroimaging are reviewed and summarized. 3D-FLAIR allows thinner slices with multi-planar reformation capability, a higher flow sensitivity, high sensitivity to subtle T1 changes in fluid, images without cerebrospinal fluid (CSF) inflow artifacts, and a 3D dataset compatible with computer-aided analysis. In addition, 3D-FLAIR can be obtained within a clinically reasonable scan time. It is important for radiologists to be familiar with the features of 3D-FLAIR and to provide useful information for patients.

Content from these authors
© 2015 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/