Single-Breath-Hold Whole-heart Unenhanced Coronary MRA Using Multi-shot Gradient Echo EPI at 3T: Comparison with Free-breathing Turbo-field-echo Coronary MRA on Healthy Volunteers
Yuji IyamaTakeshi NakauraYasunori NagayamaSeitaro OdaDaisuke UtsunomiyaMasafumi KidohHideaki YukiKenichiro HirataTomohiro NamimotoMika KitajimaKosuke MoritaYoshinori FunamaAtsushi TakemuraTomoyuki OkuakiYasuyuki Yamashita
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: mp.2017-0037

Details
Abstract

Purpose: We investigated the feasibility of single breath hold unenhanced coronary magnetic resonance angiography (MRA) using multi-shot gradient echo planar imaging (MSG-EPI) on a 3T-scanner.

Methods: Fourteen volunteers underwent single breath hold coronary MRA with a MSG-EPI and free-breathing turbo field echo (TFE) coronary MRA at 3T. The acquisition time, signal to noise ratio (SNR), and the contrast of the sequences were compared with the paired t-test. Readers evaluated the image contrast, noise, sharpness, artifacts, and the overall image quality.

Results: The acquisition time was 88.1% shorter for MSG-EPI than TFE (24.7 ± 2.5 vs 206.4 ± 23.1 sec, P < 0.01). The SNR was significantly higher on MSG-EPI than TFE scans (P < 0.01). There was no significant difference in the contrast on MSG-EPI and TFE scans (1.8 ± 0.3 vs 1.9 ± 0.3, P = 0.24). There was no significant difference in image contrast, image sharpness, and overall image quality between two scan techniques. The score of image noise and artifact were significantly higher on MSG-EPI than TFE scans (P < 0.05).

Conclusion: The single breath hold MSG-EPI sequence is a promising technique for shortening the scan time and for preserving the image quality of unenhanced whole heart coronary MRA on a 3T scanner.