Rapid MR Angiography Using 3D Gradient-echo Imaging and the Two-point Dixon Method to Evaluate Carotid Plaque
Ryusuke IrieShiori AmemiyaTsuyoshi UeyamaYuichi SuzukiHidemasa TakaoOsamu Abe
著者情報
ジャーナル オープンアクセス 早期公開

論文ID: tn.2022-0002

この記事には本公開記事があります。
詳細
抄録

Liver acquisition with volume acceleration-flex (LAVA-Flex) acquires out-of-phase and in-phase echo images and automatically generates water-only and fat-only images from one single acquisition. The scan time of carotid MR angiography (MRA) using LAVA-Flex (LAVA MRA) is about one-fifth that of conventional time-of-flight MRA (cTOF MRA). We aimed to investigate whether LAVA MRA could provide useful information for the diagnosis of carotid plaque by utilizing the ability to acquire multiple sequences simultaneously. Comparing LAVA MRA and cTOF MRA images for carotid plaque, low-intensity plaques were more clearly identified in the in-phase images, and high-intensity plaques were more clearly identified in the water-only or out-of-phase images. None of the plaques exhibited superior visualization with the cTOF sequence. We concluded that LAVA MRA can provide more useful information on plaque evaluation using multiple sequences than cTOF MRA.

著者関連情報
© 2022 by Japanese Society for Magnetic Resonance in Medicine

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.