Neurosonology:神経超音波医学
Print ISSN : 0917-074X
ISSN-L : 0917-074X
経頭蓋超音波ドプラ法の血流検出能に関する臨床的検討
永井 秀政高家 幹夫森竹 浩三
著者情報
ジャーナル フリー

1994 年 7 巻 1 号 p. 18-23

詳細
抄録
[Purpose] It has been reported that the failure rate in recording middle cerebral artery (MCA) flow ranged from 20 to 40%. In our experience, this failure rate was about 10%. The detectability of MCA by transcranial Doppler ultrasonography (TCD) is influenced by several factors: (1) temporal bone thickness. (2) blood flow velocity. (3) pulsatility index. (4) age and sex of patients. (5) branching type of the MCA. (6) hematocrit, PaCO2, systolic blood pressure of the patients. and (7) technique. In this study, we investigated the detectability in 58 vessels of 33 subjects to assess the factors that affect Doppler detectability by TCD. [Materials and Methods] We analyzed three groups classified by the detectability, namely group A (easy detection) and group B (can detect MCA flow within 10 minutes) or group C (impossible to detect) . [Results] The significant factors that influenced the measurement of MCA flow were as follows: presence of thick temporal bone, a low flow velocity, and being an elderly female. The other factors had no apparent significance due to different conditions. [Conclusion] Thickness of the skull and blood flow velocity were considered to be two major factors that make it difficult to detect intracranial blood flow by TCD.
著者関連情報
© 日本脳神経超音波学会
前の記事
feedback
Top