Orthopedics & Traumatology
Online ISSN : 1349-4333
Print ISSN : 0037-1033
ISSN-L : 0037-1033
Wear and Revision Operation of H. D. P. Artificial Femoral Head
K. HirotaH. HorikawaH. KiyaK. AkasakiY. Hayashi
Author information
JOURNAL FREE ACCESS

1976 Volume 25 Issue 3 Pages 313-316

Details
Abstract

High-Density Polythylen (H. D. P.) artificial femoral head had been applied to 35 cases of the fracture of femur neck, but many of them had brought down unsuccess within 2 years after replacement operation. By revision operation of 6 cases of them, it was confirmed that unsuccess was caused of the wear of H. D. P. head. Histology showed that fine particles of the weared H. D. P. induce the foreign body inflammation granuloma in the capsule and in the neighbor bone tissue. Bone destruction comes first to proximal region of the femur and finally to the acetabulum.

Content from these authors
© West-Japanese Society of Orthopedics & Traumatology
Previous article Next article
feedback