Optical Review
Print ISSN : 1340-6000
ISSN-L : 1340-6000
Simultaneous Measurement of Refractive Index and Thickness of Transparent Plates by Low Coherence Interferometry
Masato OHMITakehisa SHIRAISHIHideyuki TAJIRIMasamitsu HARUNA
著者情報
ジャーナル フリー

1997 年 4 巻 4 号 p. 507-515

詳細
抄録

We demonstrate a novel low coherence Michelson interferometer which can provide simultaneous measurement of the refractive index and thickness of transparent plates used as a measured object. Unlike the existing low coherence interferometers reported so far, either an object or a focusing lens aligned on the signal arm is scanned repeatedly by a precise translation stage in synchronization with movement of a reflection mirror on the reference arm. The so-called object or lens scanning method gives us two measured quantities a movement distance of the stage between two light focusing states on the front and rear planes of an object and the corresponding optical path difference. These two measured quantities, result in desirable values of the index and thickness of the object with a short calculation. The measurement accuracy of .1% is expected for a thickness of more than 1 mm. In the experiment using the object scanning method, the accuracy of 0.3% or less was successfully attained for nearly 1-mm thick plates of fused quartz, sapphire, LiTaO3 and slide glass.

著者関連情報

この記事は最新の被引用情報を取得できません。

© 1997 by the Optical Society of Japan
前の記事 次の記事
feedback
Top