STAIBILITY OF BRONCHODILATERS USING NEW ULTRASONIC NEBULIZER
Kouichiro Atsuda[in Japanese]Shigeru MotohashiSetsuko MuraseYuji Yoshiyama[in Japanese]Fumiya Tomonaga
Author information
JOURNAL FREE ACCESS

1995 Volume 38 Issue Supplement2 Pages 114-118

Details
Abstract
Ultrasonic nebulization is widely performed for inhaant therapy. But there still remain some problems in the stability of the drugs in inhalant solutions during ultrasonic nebulization. We have reported that the remaining ratios of some bronchodilaters were below 90% after nebulization using the conventional ultrasonic nebulizer.
A new ultrasonic nebulizer has been developed by OMRON. We investigated the effect of the new ultrasonic nebulizer on the stability of bronchodilaters for inhalant therapy. Stability tests were performed on bronchodilaters in terms of visible change, pH and degradation rate of the drug using HPLC after ultrasonic nebulization.
The results may be summarized as follows. The ratios of all bronchodilaters remained at more than 95% after 8 minutes of ultrasonic nebulization using the new ultrasonic nebulizer.
Therefore, bronchodilaters are highly stableusing new ultrasonic nebulizer and should be effective. We concluded that the new ultrasonic nebulizing method can be applied in the inhalant therapy of many inhalant preparations.