Preprints of Annual Meeting of The Ceramic Society of Japan
Preprints of Fall Meeting of The Ceramic Society of Japan
24th Fall Meeting of The Ceramic Society of Japan
Session ID : 3E17
Conference information

Study on measurement of temperature of the sample heated with fast heating rate by using thermo-balance with infrared heating lamp
*Katsuyoshi Oh-ishiTomoki MatsunagaRyota Kobayashi
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
We measured temperature and DTA signal of pure Li2CO3 which was heated with the heating of rate 400C/min in air by using thermo-balance with infrared heating lamp. Li2CO3 was set on thermocouple in a thermo-balance and heated with the heating of rate 400C/min up to a target temperature around the melting point of the Li2CO3 (700C, 710C, 720C)and kept at the temperature in air. During the heating process, temperature and DTA signal of the Li2CO3 sample were measured against measurement time. From theresults, DTA signal was found at the temperature above the melting point of pure Li2CO3.
Content from these authors
© The Ceramic Society of Japan 2011
Previous article Next article
feedback