Progress of Digestive Endoscopy
Online ISSN : 2187-4999
Print ISSN : 1348-9844
ISSN-L : 1348-9844
Case report
Case of patency capsule remaining in small bowel stricture after endoscopic balloon dilatio
Go KuwataSatomi KoizumiTaku TabataKoichi Koizumi
Author information
Keywords: Patency Capsule
JOURNAL FREE ACCESS

2015 Volume 87 Issue 1 Pages 162-163

Details
Abstract
A 71-year-old woman presented with recurrent abdominal pain. She was diagnosed with Crohn’s disease with ileum stenosis. She underwent endoscopic balloon dilation. At 76 years old, she again presented with lower abdominal pain and diarrhea. She underwent bowel patency testing using a patency capsule (PC) . PC was not excreted within 33 hours, and radiography localized it in the terminal ileum. After 4 days, she again experienced abdominal pain and anorexia. Radiography could not detect the PC, so that it appeared to have dissolved. Symptoms lasted for 7 days, and she underwent colonoscopy. Endoscope 9.2mm in diameter easily passed through the stricture of the ileum, but film that had coated the PC after it dissolved remained at the oral side of stricture. After the film’s removal endoscopically, symptoms resolved. PC may be caught in a small bowel stricture, so attention is required when a PC was not excreted.
Fullsize Image
Content from these authors
© 2015 Japan Gastroenterological Endoscopy Society Kanto Chapter
Previous article Next article
feedback