Progress of Digestive Endoscopy
Online ISSN : 2187-4999
Print ISSN : 1348-9844
ISSN-L : 1348-9844
症例
悪性胃流出路閉塞に対して留置した金属ステントが早期に断裂した1例
江藤 宏幸中原 守康葛西 豊高荻野 太郎村松 誠司宮原 庸介川辺 晃一
著者情報
ジャーナル フリー

2021 年 98 巻 1 号 p. 72-74

詳細
抄録

A woman in her 80s underwent placement of a self-expandable metal stent (SEMS) for malignant gastric outlet obstruction due to advanced gastric cancer. A fracture in the SEMS was observed by esophagogastroduodenoscopy and abdominal radiograph 98 days after SEMS placement, and a second SEMS was placed. She died of the primary disease 117 days after placement of the initial SEMS. Fracture of a SEMS is a relatively rare event that occurs in 0.5-0.6% of cases. Previously, it was thought to be caused by a fault in the manufacturing process or deterioration over time. In our case, when fracture of the SEMS was found, Virchow lymph node metastasis and peritoneal dissemination were also observed. External physical force due to progression of the primary lesion may cause fracture of SEMS.

著者関連情報
© 2021 一般社団法人 日本消化器内視鏡学会 関東支部
前の記事 次の記事
feedback