Plasma and Fusion Research
Online ISSN : 1880-6821
ISSN-L : 1880-6821
Regular Articles
Preliminary Design of Imaging Bolometer Fields of View for Wendelstein7-X
Byron J. PETERSONKiyofumi MUKAIRyuichi SANOHaruki KOJIMALHD Experiment GroupDaihong ZHANGMarcin JAKUBOWSKIMaciej KRYCHOWIAKYuhe FENGRalf KOENIGThomas Sunn PEDERSENthe W7-X Experiment Group
著者情報
ジャーナル フリー

2016 年 11 巻 p. 2402101

詳細
抄録

The InfraRed imaging Video Bolometer (IRVB) provides absolutely calibrated images of the plasma radiation. In this paper we present the design of four imaging bolometers viewing divertor and non divertor regions of Wendelstein7-X (W7-X). This design assumes a 640 × 480 pixel, 30 fps, 100 mK, IR camera imaging each IRVB foil. The bolometer cameras consist of a 9 cm × 7 cm × 2 µm thick Pt foils blackened with graphite, which are located distances ranging 6cm to 8cm from a 4.3 mm × 4.3 mm aperture. The 8 cm × 6 cm central section of the foil is divided into 24 × 18 IRVB channels each consisting of 504 IR camera pixels. Using a previously derived expression, the noise equivalent power of the IRVB is calculated as 109 µW/cm2. Assuming 5 MW of radiated power uniformly emanating from the 30 m3 plasma, the signal level is estimated as ranging from 4.5 to 8.1 mW/cm2 giving signal to noise ratios between 42 and 74.

著者関連情報
© 2016 by The Japan Society of Plasma Science and Nuclear Fusion Research
前の記事 次の記事
feedback
Top