Plasma and Fusion Research
Online ISSN : 1880-6821
ISSN-L : 1880-6821
Regular Articles
Development of a Wide Dynamic Range Neutron Flux Measurement Instrument Having Fast Time Response for Fusion Experiments
Daijiro ITOHiroyuki YAZAWAMakoto TOMITAKATsuyoshi KUMAGAIShigehiro KONOMichinori YAMAUCHITsuyoshi MISAWATakashi KOBUCHIHiroshi HAYASHIHitoshi MIYAKEKunihiro OGAWATakeo NISHITANIMitsutaka ISOBE
Author information
JOURNAL FREE ACCESS

2021 Volume 16 Pages 1405018

Details
Abstract

A wide-range neutron flux measurement instrument is developed herein for monitoring the total neutron emission rate and yield of the Large Helical Device (LHD) during deuterium experiments implemented from March 2017 in the National Institute for Fusion Science (NIFS), Japan. The instrument is designed for and installed on the Neutron Flux Monitoring (NFM) system, which measures the counting rate using a 235U Fission Chamber. By combining the pulse counting and Campbell methods, the Digital Signal Processing Unit (DSPU) realized a wide dynamic range of over six orders of magnitude from 1 × 103 counts/s (cps) to 5 × 109 cps. This study explains and discusses how the instrument is developed, including topics from the predevelopment activities to the verification test at the Kyoto University Critical Assembly (KUCA). Experimental results in the LHD using the finished products suggest that the NFM system works well during deuterium experiments.

Content from these authors
© 2021 by The Japan Society of Plasma Science and Nuclear Fusion Research
Previous article Next article
feedback