Radiocarbon Measurement of Purified Bio Diesel Fuel by a Liquid Scintillation Counter
Masaaki SAITO
Author information
JOURNALS OPEN ACCESS

2009 Volume 58 Issue 7 Pages 455-460

Details
Abstract

Bio diesel fuel(BDF) was purified by column chromatography and the carbon 14(C-14) content was successfully measured by liquid scintillation counting(LSC). BDF was dissolved in hexane (10 wt%), and was purified by the column chromatography using an activated clay. Then the purified BDF was separated from hexane by evaporation under the boiling point of hexane. The purified BDF with toluene scintillator was measured by LSC. The biomass ratio obtained by LSC was in good agreement with that obtained by accelerator mass spectrometry.