Report on Present Status of Quality Control and Assurance of PET in Japan Based on Questionnaire Survey
Subcommittee for Radionuclide Imaging and Nuclear Medicine Technology, Medical and Pharmaceutical Committee, Japan Radioisotope Association Medical and Pharmaceutical Committee, Japan Radioisotope Association
Author information
JOURNAL OPEN ACCESS

2012 Volume 61 Issue 11 Pages 555-569

Details
Abstract
The Japan Radioisotope Association conducted a questionnaire survey to ascertain status of QA/QC activity of PET in Japan. The questionnaire was sent to 277 PET facilities with 116 valid responses(41.9%). The survey revealed that acceptance test of PET was performed by vender in 86.0% and by facility in 7.6% of the scanners. System sensitivity, accuracy of image registration, image homogeneity, accuracy of dead-time correction and accidental coincidence, and accuracy of attenuation and scatter correction was tested in 59.6%, 57.9%, 55.0%, 29.2%, and 28.1% of the PET scanners, respectively. The count normalization and cross calibration was tested once in 3 months in 71.5% of the scanners. Maintenance service contract was present in 93.3%. Normalization factor, cross calibration factor, and SUV were varied from −11 to +20%, from −8 to +54%, and from −20 to +10% between the initial and the last tests. Dose calibrator and well counter was regularly checked in 55.0% and 82.6% of the equipment, respectively.