1965 Volume 14 Issue 2 Pages 73-77
The present paper describes a new chemical dosimeter utilizing K2Cr2O7⋅H2SO4 system. Decrease of extinction of γ-irradiated K2Cr2O7 were measured in the wavelength of 435 and 350 millimicron. The results obtained are as follows;
1. Decrease of extinction is proportional to radiation dose and independent of dose rate.
2. G-Value is 0.446±0.023.
3. In the dose range of 104r-107r, properly prepared solution of this system can offer reliable results.
4. The reactions in this system are relatively stable, so the total dose in the divided irradiation is also measurable.