A Thin 4πβ-Counter Operating by Negative High Voltage for the 4πβ-γ Coincidence Measurement
-Operating Characteristics-
Hiroshi MIYAHARAMasakazu SUZUKITamaki WATANABE
Author information
JOURNAL FREE ACCESS

1978 Volume 27 Issue 11 Pages 625-630

Details
Abstract

A thin 4πβ-counter with five anode wires was operated by supplying negative high voltage to the inner wall and showed good plateau characteristics in the case of using only the center anode to take out signals in the previous experiments. This counter, as the 4πβ-counter and both 2πβ-counters, showed good plateau characteristics in these detailed experiments and no dependency on the disintegration rates examined up to 6000 dps. Then a 4πβ-γ coincidence equipment was set up using this 4πβ-counter and one NaI (Tl) scintillation detector with size of 76 mmφ×76 mm. The radioactivities that were measured for β-γ and electron capture-γ decay nuclides by this coincidence equipment were coincided with those measured by the other one. The improvement of the γ-ray detection efficiency to aim was achieved by that the efficiency obtained by this system was the same as one by the other system having two NaI (Tl) detectors with the same size.