Optimization of Liquid Scintillation Counting by Using Multichannel Pulse-Height Analyzer
Nobuhiko ITOMasamichi IKEDA
Author information
JOURNAL FREE ACCESS

1984 Volume 33 Issue 8 Pages 539-542

Details
Abstract
The method for determining optimum conditions of a liquid scintillation counter (LSC) for lowlevel counting of tritium was studied. To compare the lower limit of detection, the figure of merit EM/√B was used. The measurement system is a LSC connected with a multichannel pulse-height analyzer (MCA) . Both of tritium spectra and background spectra were searched for the condition that maximizes the figure of merit. The optimum condition obtained by using the MCA was fitted for a single channel pulse-height analyzer of the LSC. The figure of merit obtained by this method was 20.6% greater than that obtained by the balance point method.