Leakage Radiations in a Medical Electron Accelerator Facility
—Measurement of Neutrons and Comparison with Calculation—
Kaneo MORIKAWAKenji ISHIMATSU
Author information
JOURNAL FREE ACCESS

1990 Volume 39 Issue 2 Pages 37-44

Details
Abstract

Neutron dose equivalent rates are measured in the irradiation room and the maze of a medical electron accelerator facility. Are m-counter used for measurement is associated with a gate circuit to reject piled up signals caused by pulsed X-rays, so that it is possible to measure the neutrons without obstruction by the piled up signals even in the irradiation room. Counting loss of neutrons is mathematically corrected, and the correction factor comes nearly equal to 1.5 at dose points close to the entrance to the irradiation room when 14 MV X-rays are generated. The measurement in the maze agrees with the calculation reported in the preceding report in a factor less than 1.5 except for dose points close to the entrance to the irradiation room.