Radiation Safety Management
Online ISSN : 1884-9520
Print ISSN : 1347-1511
ISSN-L : 1347-1511
Papers
Response of a Radiophotoluminescence Glass Dosimeter to Low Energy X-rays Necessary for Mammography
Daisuke MakiHatsumi KobayashiSatoru HisakadoFuminobu SatoYushi KatoTakayoshi YamamotoToshiyuki Iida
著者情報
ジャーナル フリー

2009 年 8 巻 1 号 p. 8-13

詳細
抄録
  New-type radiophotoluminescence (RPL) glass dosimeters, which had thin sensitive layers near their surface, were made for the examination of their response to low energy (15-25 keV) X-rays necessary for mammography. We constructed a laser scanning microscope system for the effective measurement of RPL photons from an X-ray irradiated glass dosimeter. The X-ray sensitive region of the new-type glass dosimeter was near its surface, and moreover a thin X-ray absorption glass filter, whose photon attenuation coefficient was almost the same as the glass dosimeter, was put on the new-type glass dosimeter. These new-type glass dosimeters were irradiated with low energy X-rays and measured results of absorbed dose around their surface layers agreed on the whole with results from photon transport calculations. It was confirmed that RPL glass dosimeters could make a reliable service to the low energy X-ray dosimetry for mammography.
著者関連情報
© 2009 Japanese Society of Radiation Safety Managenent
前の記事 次の記事
feedback
Top