金属表面技術
Online ISSN : 1884-3395
Print ISSN : 0026-0614
ISSN-L : 0026-0614
分光光度計を用いたアルミニウム表面処理皮膜の膜厚測定
内山 利光大塚 達雄北村 照夫多田 清志
著者情報
ジャーナル フリー

1986 年 37 巻 2 号 p. 78-81

詳細
抄録

Recently, the need has increasingly been felt for a non-destructive, rapid and precise method of testing, which can be used to measure the thickness of functional film. Use of the microscope and eddy-current instrument recommended by JIS H8680, however, produces unsatisfactory test results for thicknesses below 2μm.
Thickness of anodic oxide films and chromate coatings was therefore measured using interference fringes obtained by spectrophotometer, and the results obtained were compared with those obtained by microscope and TEM in terms of precision and measurement limits. The thickness test results obtained for sulfuric film and chromate coating using the spectrophotometer were good, with mimits for the thickness test of 0.14μm on sulfuric film (n=1.6) and 0.11μm on chromate coating (n=2.0) at wavelengths from 200 to 2500nm, where n is the refractive index.

著者関連情報
© (社)表面技術協会
前の記事 次の記事
feedback
Top