A new SF6 inlet system with a modified Faraday collector alignment of Finnigan MAT-251 mass spectrometer for sulfur isotope measurement
Madhusoodhan Satish-KumarJunshi MatsudaRie YamazakiAoi TakanoHideki Wada
著者情報
研究報告書・技術報告書 フリー

2010 年 37 巻 p. 41-49

詳細
抄録

A new inlet system for the preparation of SF6 gas by fluorination of sulfide minerals, suitable for both in-situ laser ablation and powder samples, was connected to the existing CO2 inlet system of the Finnigan MAT-251 mass spectrometer at Shizuoka University. The preparation line is composed of a fluorine generator, two types of reaction chambers, capacitance manometer, KBr canister and liquid nitrogen slush connected by pneumatic bellows valves. It also includes a working standard SF6 gas reservoir and a machine standard SF6 gas reservoir for routine analyses. The preparation line can hold high vacuum conditions without appreciable leaks for intervals more than that required to complete one sample measurement. For simultaneous measurement of [127], [128], [129] and [131] isotopes of SF5+, we have renovated Faraday collectors by reducing its body thickness from 4.5 mm to 3.5 mm and realigned the collector configuration of the mass spectrometer. Preliminary results of δ34S values for working standard SF6 gas gave high precession of ±0.003‰ . Further modifications in the collector assembly are in progress for the simultaneous measurement of the closely spaced ion trajectory of [127], [128] and [129] isotopes of SF5+.