トルク計測機器の静校正における不確かさの実験的検証
第1報 —繰り返し性、設置変更の再現性及び零点誤差の検証—
大串 浩司太田 孝上田 和永
著者情報
会議録・要旨集 フリー

p. 27

詳細
抄録
As uncertainty contributions for the static torque calibration of a torque measuring device, the following components are considered; repeatability with unchanged mounting position, reproducibility with changed mounting position, zero error, interpolation error, reversibility (hysteresis) and resolution including the ability of the indicator. Some foreign standard documentations, however, have not revealed any experimental confirmation for the validity of assumed probability distribution for each contribution. In order to establish a determination method of uncertainty for the torque calibration, the validities of probability distributions were verified by the numerous repetitions of torque measurement using the primary torque standard machine. The determination method is proposed based on the experimental results.